Programming the Status Register System

Status Groups

Data Questionable Condition Register

The Data Questionable Condition Register continuously monitors the hardware and firmware status of the signal generator. Condition registers are read only.

Table 4-7 Data Questionable Condition Register Bits

Bit

Description

 

 

0, 1, 2

Unused. These bits are always set to 0.

 

 

3

Power (summary). This is a summary bit taken from the QUEStionable:POWer register. A 1 in this bit position

 

indicates that one of the following may have happened: The ALC (Automatic Leveling Control) is unable to

 

maintain a leveled RF output power (i.e., ALC is UNLEVELED), the reverse power protection circuit has been

 

tripped. See the “Data Questionable Power Status Group” on page 168 for more information.

 

 

4

N5181A/82A/83A: ALC Heater Detector (COLD). A 1 in this bit position indicates that the ALC detector is

 

cold.

 

E4428C/38C, E8257D/67D, and E8663B: Temperature (OVEN COLD). A 1 in this bit position indicates that

 

the internal reference oscillator (reference oven) is cold.

 

 

5

Frequency (summary). This is a summary bit taken from the QUEStionable:FREQuency register. A 1 in this bit

 

position indicates that one of the following may have happened: synthesizer PLL unlocked, 10 MHz reference

 

VCO PLL unlocked, 1 GHz reference unlocked, sampler, YO loop unlocked or baseband 1 unlocked. For more

 

information, see the “Data Questionable Frequency Status Group” on page 171.

 

 

6

Unused. This bit is always set to 0.

 

 

7

Modulation (summary). This is a summary bit taken from the QUEStionable:MODulation register. A 1 in this

 

bit position indicates that one of the following may have happened: modulation source 1 underrange,

 

modulation source 1 overrange, modulation source 2 underrange, modulation source 2 overrange, or modulation

 

uncalibrated. See the “Data Questionable Modulation Status Group” on page 174 for more information.

 

 

8ab

Calibration (summary). This is a summary bit taken from the QUEStionable:CALibration register. A 1 in this

 

bit position indicates that one of the following may have happened: an error has occurred in the DCFM/DCΦM

 

zero calibration, or an error has occurred in the I/Q calibration. See the “Data Questionable Calibration Status

 

Group” on page 177 for more information.

 

 

9

Self Test. A 1 in this bit position indicates that a self- test has failed during power- up. Reset this bit by cycling

 

the signal generator’s line power. *CLS will not clear this bit.

 

 

10, 11

Unused. These bits are always set to 0.

 

 

12c

BERT (summary). This is a summary bit taken from the QUEStionable:BERT register. A 1 in this bit position

 

indicates that one of the following occurred: no BCH/TCH synchronization, no data change, no clock input,

 

PRBS not synchronized, demod/DSP unlocked, or demod unleveled. See the “Data Questionable BERT Status

 

Group” on page 180 for more information.

 

 

13, 14

Unused. These bits are always set to 0.

 

 

15

Always 0.

 

 

a.In the N5182A, this bit applies only to the I/Q calibration. In the N5181A/83A, this bit is unused and always set to 0. b.In the E8257D, and the E8663B this bit applies only to the DCFM/DCΦM calibration.

c.In the N5181A, N5182A, N5183A, E4428C, E8257D, E8267D, and the E8663B this bit is always set to 0.

Query: STATus:QUEStionable:CONDition?

166

Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

Page 176
Image 176
Agilent Technologies N5181A/82A, N5183A MXG, E8663B manual Data Questionable Condition Register Bits

N5183A, N5183A MXG, E8663B, N5181A/82A specifications

Agilent Technologies is renowned for its innovative solutions in electronic test and measurement equipment. Among its offerings are several signal generators including the N5181A, N5182A, E8663B, and N5183A MXG. These models are distinguished not only by their performance but also by their versatility across various applications in communications, aerospace, and electronics.

The Agilent N5181A and N5182A, part of the MXG family, are highly versatile signal generators known for their exceptional frequency performance and flexibility. The N5181A operates from 100 kHz to 6 GHz, while the N5182A extends that range up to 12 GHz. They provide high fidelity signals with low phase noise, making them ideal for the development and testing of RF components and systems. These generators support a wide variety of modulation formats, including AM, FM, PM, and pulse modulation, catering to diverse application needs.

The E8663B, meanwhile, is designed for the range of 250 kHz to 3 GHz and is also recognized for its high-performance features. Its built-in capabilities for modulation make it effective for testing wireless devices, ensuring that signals can be simulated accurately in both laboratory and field environments. It is especially beneficial for users requiring a straightforward and efficient solution with high reliability.

The N5183A MXG signal generator enhances the lineup with frequency coverage up to 6 GHz and advanced capabilities. It integrates various modulation capabilities while ensuring high signal integrity. Its architecture is tailored for both production test environments and research applications, providing users with the flexibility to adapt to changing testing requirements.

Common characteristics across these models include a user-friendly interface that simplifies configuration and operation. They are often equipped with LAN and USB interfaces for easy remote control and integration into automated test systems. The robustness of these generators allows them to perform reliably in challenging environments, making them essential tools in laboratories, manufacturing floors, and field testing scenarios.

In summary, Agilent's signal generators, including the N5181A, N5182A, E8663B, and N5183A MXG, represent a blend of advanced technology, flexibility, and precision. These instruments are vital in facilitating the evolution of cutting-edge communication technologies, ensuring that designers and engineers can confidently meet the demands of modern electronics.