Programming the Status Register System

Status Groups

Standard Event Status Register

Table 4-4

Standard Event Status Register Bits

 

 

 

 

Bit

 

Description

 

 

 

 

0

 

Operation Complete. A 1 in this bit position indicates that all pending signal generator operations were completed

 

 

following execution of the *OPC command.

 

 

 

1

 

Request Control. This bit is always set to 0. (The signal generator does not request control.)

 

 

 

2

 

Query Error. A 1 in this bit position indicates that a query error has occurred. Query errors have instrument error

 

 

numbers from −499 to −400.

 

 

 

3

 

Device Dependent Error. A 1 in this bit position indicates that a device dependent error has occurred. Device

 

 

dependent errors have instrument error numbers from −399 to −300 and 1 to 32767.

 

 

 

4

 

Execution Error. A 1 in this bit position indicates that an execution error has occurred. Execution errors have

 

 

instrument error numbers from −299 to −200.

 

 

 

5

 

Command Error. A 1 in this bit position indicates that a command error has occurred. Command errors have

 

 

instrument error numbers from −199 to −100.

 

 

 

6

 

User Request Key (Local). A 1 in this bit position indicates that the Local key has been pressed. This is true even if

 

 

the signal generator is in local lockout mode.

 

 

 

7

 

Power On. A 1 in this bit position indicates that the signal generator has been turned off and then on.

 

 

 

 

 

 

Query:

*ESR?

 

 

Response:

The decimal sum of the bits set to 1

 

 

Example:

The decimal value 136 is returned. The decimal sum = 128 (bit 7) + 8 (bit 3).

Standard Event Status Enable Register

The Standard Event Status Enable Register lets you choose which bits in the Standard Event Status Register set the summary bit (bit 5 of the Status Byte Register) to 1.

*ESE <data>

<data> is the sum of the decimal values of the bits you want to enable.

Example:

To enable bit 7 and bit 6 so that whenever either of those bits are set to 1, the Standard Event

 

Status summary bit of the Status Byte Register is set to 1. Send the command *ESE 192 (128 +

 

64).

Query:

*ESE?

Response:

Decimal value of the sum of the bits previously enabled with the *ESE <data> command.

158

Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

Page 168
Image 168
Agilent Technologies N5181A/82A, N5183A MXG manual Standard Event Status Register, Standard Event Status Enable Register

N5183A, N5183A MXG, E8663B, N5181A/82A specifications

Agilent Technologies is renowned for its innovative solutions in electronic test and measurement equipment. Among its offerings are several signal generators including the N5181A, N5182A, E8663B, and N5183A MXG. These models are distinguished not only by their performance but also by their versatility across various applications in communications, aerospace, and electronics.

The Agilent N5181A and N5182A, part of the MXG family, are highly versatile signal generators known for their exceptional frequency performance and flexibility. The N5181A operates from 100 kHz to 6 GHz, while the N5182A extends that range up to 12 GHz. They provide high fidelity signals with low phase noise, making them ideal for the development and testing of RF components and systems. These generators support a wide variety of modulation formats, including AM, FM, PM, and pulse modulation, catering to diverse application needs.

The E8663B, meanwhile, is designed for the range of 250 kHz to 3 GHz and is also recognized for its high-performance features. Its built-in capabilities for modulation make it effective for testing wireless devices, ensuring that signals can be simulated accurately in both laboratory and field environments. It is especially beneficial for users requiring a straightforward and efficient solution with high reliability.

The N5183A MXG signal generator enhances the lineup with frequency coverage up to 6 GHz and advanced capabilities. It integrates various modulation capabilities while ensuring high signal integrity. Its architecture is tailored for both production test environments and research applications, providing users with the flexibility to adapt to changing testing requirements.

Common characteristics across these models include a user-friendly interface that simplifies configuration and operation. They are often equipped with LAN and USB interfaces for easy remote control and integration into automated test systems. The robustness of these generators allows them to perform reliably in challenging environments, making them essential tools in laboratories, manufacturing floors, and field testing scenarios.

In summary, Agilent's signal generators, including the N5181A, N5182A, E8663B, and N5183A MXG, represent a blend of advanced technology, flexibility, and precision. These instruments are vital in facilitating the evolution of cutting-edge communication technologies, ensuring that designers and engineers can confidently meet the demands of modern electronics.