Features and Functions

2

The major functional blocks of the U2802A module are:

Analog input channel circuitry

Cold junction sensor

Digital control logic

EEPROM

Analog input channel circuitry

The analog circuitry for each channel consists of an instrumentation amplifier with a fixed gain of 97.673, a 4 Hz RC low- pass filter, and an output buffer. The multiplexers at the input and output of each channel allows each channel to be configured for three modes of operation as listed below:

Thermocouple input mode: In thermocouple mode, the thermocouples (or any floating voltage source) should be connected to the TCn+ and TCnterminals as illustrated in Figure 2- 4. All TCnterminals are internally tied to module ground with a 10 MΩ resistor. The TCn+ and TCnsignals are routed to the differential inputs of the instrumentation amplifier. Differential voltage signals at the TCn+ and TCnterminals are amplified, filtered and driven out by single- ended output voltage to the corresponding AI channel on Rear Connector 1.

Figure 2-4Functional block diagram for thermocouple mode in U2802A

Agilent U2802A 31-Channel Thermocouple Input User’s Guide

31

Page 47
Image 47
Agilent Technologies manual Analog input channel circuitry, 4Functional block diagram for thermocouple mode in U2802A