4 Product Specifications

Calculating System Accuracy

The overall measurement system comprises of three major components:

1DAQ (U2355A or U2356A)

2Signal Conditioner (U2802A), which includes CJ Sensor error

3Sensor (Thermocouples)

Errors introduced by each of the above components has to be accounted for when calculating the total system accuracy. Since errors from each component are not correlated with each other, the total system error will be the root- sum- square (RSS) of all the errors:

ETOTAL2 = EDAQ2 + ESIG_COND2 + ETHERMOCOUPLE2

Example:

Assume the following conditions:

DAQ: Agilent U2355A

Signal Conditioning: Agilent U2802A

Ambient temperature: 23 °C

Thermocouple type: J- type, standard limits of error

Temperature to measure: 600 °C

Assume the following error specifications:

U2355A: Gain error = 0.02% of reading

Offset error = 1 mV

U2802A gain = 97.673

Gain error = 0.06% of reading

Offset error = 15 µV (with respect to input)

Zeroing error = 6 µV (with respect to input)

CJ measurement accuracy = 1 °C

Thermocouple = greater than 2.2 °C or 0.75% error

Noise error has been omitted to simplify the example