Image 68

4 Product Specifications

 

 

 

 

Thermocouple Measurements Accuracy

 

 

 

 

 

(U2355A @ 0 °C to 18 °C and 28 °C to 45 °C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ITS-90

 

Optimum

Without

 

50 points

500 points

 

T/C

Temperature

 

Measurement

 

 

 

averaging

 

averaging

averaging

 

Type

Range (°C)

 

Range(°C)

 

 

 

(± °C)

 

(± °C)

(± °C)

 

 

 

 

 

 

 

 

 

 

Low

High

 

Low

High

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B

0

1820

 

1100

1820

3.4

 

2.4

2.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

400

1100

7.5

 

3.6

2.2

 

 

 

 

 

 

 

 

 

 

 

 

E

–270

1000

 

–150

1000

2.7

 

2.6

2.5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–150

3.8

 

3.6

3.6

 

 

 

 

 

 

 

 

 

 

 

 

J

–210

1200

 

–150

1200

2.5

 

2.4

2.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–210

–150

4.2

 

4.0

3.9

 

 

 

 

 

 

 

 

 

 

 

 

K

–270

1372

 

–100

1200

2.9

 

2.8

2.8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–100

4.3

 

4.0

3.9

 

 

 

 

 

 

 

 

 

 

 

 

N

–270

1300

 

–100

1300

2.6

 

2.5

2.5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–100

4.9

 

4.2

4.0

 

 

 

 

 

 

 

 

 

 

 

 

R

–50

1768

 

300

1760

3.8

 

3.1

3.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–50

300

8.5

 

4.6

3.3

 

 

 

 

 

 

 

 

 

 

 

 

S

–50

1768

 

400

1760

4.2

 

3.4

3.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–50

400

7.7

 

4.2

3.1

 

 

 

 

 

 

 

 

 

 

 

 

T

–270

400

 

–100

400

2.4

 

2.2

2.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–100

4.3

 

4.0

3.9

 

 

 

 

 

 

 

 

 

Table 4-2

U2802A measurement accuracy with U2355A, at 0 to 18 °C and 28 to 45 °C,

with different number of averaging points.