Agilent Technologies U2802A System Accuracy Specifications, N O Te, Product Specifications, High

Models: U2802A

1 81
Download 81 pages 3.27 Kb
Page 67
Image 67

Product Specifications

4

System Accuracy Specifications

The Agilent U2802A thermocouple input measurement accuracy with the U2355A and U2356A is as shown in Table 4- 1, Table 4- 2, and Table 4- 3.

N O TE

Assume a ±1 °C error in the CJ measurement due to sensor error and temperature

gradient error in the accuracy numbers in Table 4-1, Table 4-2, and Table 4-3 below.

 

Table 4-1, Table 4-2, and Table 4-3 are derived from the U2802A and DAQ input accuracy

 

specifications without including the thermocouple error. Refer to “Calculating System

 

Accuracy” on page 54 for calculation methodology.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Thermocouple Measurements Accuracy

 

 

 

 

 

 

 

 

(U2355A, U2356A @ 23 °C ± 5°C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ITS-90

 

Optimum

Without

 

50 points

500 points

 

T/C

Temperature

 

Measurement

 

 

 

averaging

 

averaging

averaging

 

Type

Range (°C)

 

Range(°C)

 

 

 

(± °C)

 

(± °C)

(± °C)

 

 

 

 

 

 

 

 

 

 

Low

High

 

Low

High

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B

0

1820

 

1100

1820

1.9

 

1.2

1.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

400

1100

4.4

 

2.5

2.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E

–270

1000

 

–150

1000

1.7

 

1.6

1.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–150

2.4

 

2.3

2.3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

J

–210

1200

 

–150

1200

1.6

 

1.5

1.5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–210

–150

2.7

 

2.6

2.5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

K

–270

1372

 

–100

1200

1.5

 

1.4

1.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–100

2.7

 

2.6

2.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N

–270

1300

 

–100

1300

1.5

 

1.3

1.3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–100

3.0

 

2.7

2.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R

–50

1768

 

300

1760

2.0

 

1.4

1.3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–50

300

5.0

 

3.1

2.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S

–50

1768

 

400

1760

2.1

 

1.6

1.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–50

400

4.5

 

2.8

2.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

T

–270

400

 

–100

400

1.5

 

1.4

1.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

–200

–100

2.7

 

2.5

2.5

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 4-1U2802A measurement accuracy with U2355A or U2356A, at 23 °C ± 5 °C, with different number of averaging points.