OPERATING INSTRUCTIONS
CALIBRATE PRESSURE TRANSMITTERS
READ PRESSURE AND
P-XMTR  mA, P-XMTR  %, P-XMTR  DP%
Choose this function to simultaneously supply power to a 
1)Disconnect both the pressure input and output wires from the 
2)Select the proper module for the pressure range of the process
3)Place the QuikCal 90 Module in the ModPak 91 holder
4)Plug the ModPak cable into the mating connector on the 
5)Turn the selector knob to PRESSURE
6)The 
7)The 
8)Press the °C/°F/ZERO 
9)Press the DISPLAY/SOURCE/READ 
10)Press READ/POWER TRANSMITTER to switch between reading milliamps and supply voltage to power the transmitter
11)Carefully make all pressure connections
12)Connect the red POWER lead of the 
Use the hand pump or regulated pressure source to adjust the pressure
The TechChek 
PROCESS CALIBRATOR
34  | 35  |