OPERATING INSTRUCTIONS
CALIBRATE FREQUENCY INPUTS
Choose this function to provide pulses into frequency measuring instruments. The
10.00kHz, 1 to 1000 Hz and from 1 to 1000 CPM
CPM is used to simulate extremely slow frequency signals with greater resolution. For example, 10 Hz is equivalent to 600 CPM. To convert from CPM to Hz Divide by 60. To convert from Hz to CPM multiply by 60.
1)Disconnect one or both input wires from the device to be calibrated
2)Press the FREQUENCY/PRESSURE pushbutton until any frequency signal (KHz, Hz or CPM) appears on the display.
3)Press the DISPLAY/SOURCE/READ
4)Press the TYPE/ENG UNITS pushbuuton to select between KHz, Hz or CPM on the display
5)Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus
Output frequency is continuously adjustable with the UP/DOWN
Frequency Receiver Input
Flowmeter
Controller
Transmitter
Computer
Logger
DCS
OPERATING INSTRUCTIONS
READ FUNCTIONS
Select read by pressing the DISPLAY/SOURCE/READ pushbutton until the word READ appears on the LCD display. The READ functions measure the desired signal. Multiple scales are available for some functions.
MIN/MAX
To read the Maximum or Minimum INPUT since READ mode was entered, simply press the MAX or MIN
RESTARTING MIN/MAX
Pressing the STORE/RESET
OUT OF RANGE SIGNALS
OVER | Signals above or below those available for | |
the currently selected range will be indicat- | ||
UNDER | ||
ed by OVER and UNDER on the display. |
18 | 19 |