OPERATING INSTRUCTIONS
SOURCE MODE
Select source by pressing the DISPLAY/SOURCE/READ pushbutton until the word SOURCE appears on the LCD display. To change the output value, press the UP or DOWN key for each digit. To ramp the output press and hold any UP or DOWN pushbutton. The display will continue to change in increments corresponding to the digit being changed and will automatically carry up and down until the limits of the range are reached. This function operates in all three output positions (HI, SET & LO).
STORING QUIK-CHEK OUTPUTS
1) Press HI or LO
2) Press the UP/DOWN keys to desired value
3) Press the STORE
The LCD will flash once to show that the value was saved
If a value is in the SET position and you want that value stored in HI or LO, press and hold the STORE
Then release both
RECALLING QUIK-CHEK OUTPUTS
When you need a stored value just press a
OPERATING INSTRUCTIONS
CALIBRATE MILLIAMP INPUTS
mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow)
Choose this function to provide an output from 0.00 to 24.00 milliamps. The compliance voltage is a nominal 24 VDC to provide the driving power to your milliamp receivers.
1)Disconnect one or both input wires from the device to be calibrated.
2)Press the MILLIAMP/V
3)Press the DISPLAY/SOURCE/READ
4)If
5)Press the mA/%/% DP FLOW
6)Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus
Output current is continuously adjustable with the UP/DOWN
Milliamp Receiver Input
Controller
Transmitter
Computer
Logger
I/P
DCS
8 | 9 |