THERMOCOUPLE SPECIFICATIONS
T/C  | °C  | 
  | °F  | 
  | 
TYPE | RANGE ACCURACY | RANGE  | ACCURACY  | |
  | 
  | 
  | 
  | 
  | 
J  | 100 to 1200  | ±1.7  | 212 to 2192  | ±2.3  | 
  | ±1.8  | ±2.5  | ||
  | ±2.2  | ±3.2  | ||
  | ±2.8  | ±4.3  | ||
K  | 1100 to 1372  | ±2.2  | 2012 to 2500  | ±3.1  | 
  | 0 to 1099  | ±2.0  | 32 to 2011  | ±2.9  | 
  | ±2.3  | ±3.3  | ||
  | ±3.6  | ±5.7  | ||
  | 
  | 
  | 
  | 
  | 
T  | 200 to 400  | ±1.7  | 392 to 752  | ±2.4  | 
  | 0 to 199  | ±1.9  | 32 to 391  | ±2.7  | 
  | ±2.4  | ±3.5  | ||
  | ±3.5  | ±5.6  | ||
  | 
  | 
  | 
  | 
  | 
E  | 250 to 1000  | ±1.5  | 482 to 1832  | ±2.0  | 
  | 50 to 249  | ±1.6  | 122 to 481  | ±2.1  | 
  | ±1.9  | ±2.6  | ||
  | ±2.6  | ±3.9  | ||
  | 
  | 
  | 
  | 
  | 
THERMOCOUPLE SPECIFICATIONS
T/C  | °C  | 
  | °F  | 
  | |
TYPE  | RANGE ACCURACY  | RANGE  | ACCURACY  | ||
  | 
  | 
  | 
  | 
  | |
N  | 300 to 1300  | ±2.1  | 572 to 2372  | ±3.0  | |
  | 100 to 299  | ±2.3  | 212 to 571  | ±3.3  | |
  | ±2.6  | ±4.0  | |||
  | ±5.0  | ±8.2  | |||
R  | 1750 to 1768  | ±4.2  | 3182 to 3214  | ±6.9  | |
950 to 1749  | ±4.0  | 1742 to 3181  | ±6.5  | ||
  | |||||
  | 650 to 949  | ±4.4  | 1202 to 1741  | ±7.2  | |
  | 300 to 649  | ±5.1  | 572 to 1201  | ±8.4  | |
  | 
  | 
  | 
  | 
  | |
S  | 1700 to 1768  | ±4.9  | 3092 to 3214  | ±8.0  | |
1050 to 1699  | ±4.4  | 1922 to 3091  | ±7.1  | ||
  | |||||
  | 700 to 1049  | ±4.8  | 1292 to 1921  | ±7.8  | |
  | 300 to 699  | ±5.4  | 572 to 1291  | ±8.9  | |
  | 
  | 
  | 
  | 
  | |
Note: Thermocouple accuracies are based on an 80.00 mV Span T/C Accuracy for °C is ±(0.05% of 80.00 mV + 1°C)
T/C Accuracy for °F is ±(0.05% of 80.00 mV + 1°F)
Source resolution is 1 °C or °F. Read resolution is 0.1 °C or °F
40  | 41  |