THERMOCOUPLE SPECIFICATIONS

T/C

°C

 

°F

 

TYPE

RANGE ACCURACY

RANGE

ACCURACY

 

 

 

 

 

J

100 to 1200

±1.7

212 to 2192

±2.3

 

-50 to 99

±1.8

-58 to 211

±2.5

 

-150 to -49

±2.2

-238 to -57

±3.2

 

-200 to -149

±2.8

-328 to -237

±4.3

K

1100 to 1372

±2.2

2012 to 2500

±3.1

 

0 to 1099

±2.0

32 to 2011

±2.9

 

-100 to -1

±2.3

-148 to 31

±3.3

 

-200 to -99

±3.6

-328 to -147

±5.7

 

 

 

 

 

T

200 to 400

±1.7

392 to 752

±2.4

 

0 to 199

±1.9

32 to 391

±2.7

 

-100 to -1

±2.4

-148 to 31

±3.5

 

-200 to -99

±3.5

-328 to -147

±5.6

 

 

 

 

 

E

250 to 1000

±1.5

482 to 1832

±2.0

 

50 to 249

±1.6

122 to 481

±2.1

 

-100 to 49

±1.9

-148 to 121

±2.6

 

-200 to -99

±2.6

-328 to -147

±3.9

 

 

 

 

 

THERMOCOUPLE SPECIFICATIONS

T/C

°C

 

°F

 

TYPE

RANGE ACCURACY

RANGE

ACCURACY

 

 

 

 

 

N

300 to 1300

±2.1

572 to 2372

±3.0

 

100 to 299

±2.3

212 to 571

±3.3

 

-50 to 99

±2.6

-58 to 211

±4.0

 

-200 to -49

±5.0

-328 to -57

±8.2

R

1750 to 1768

±4.2

3182 to 3214

±6.9

950 to 1749

±4.0

1742 to 3181

±6.5

 

 

650 to 949

±4.4

1202 to 1741

±7.2

 

300 to 649

±5.1

572 to 1201

±8.4

 

 

 

 

 

S

1700 to 1768

±4.9

3092 to 3214

±8.0

1050 to 1699

±4.4

1922 to 3091

±7.1

 

 

700 to 1049

±4.8

1292 to 1921

±7.8

 

300 to 699

±5.4

572 to 1291

±8.9

 

 

 

 

 

Note: Thermocouple accuracies are based on an 80.00 mV Span T/C Accuracy for °C is ±(0.05% of 80.00 mV + 1°C)

T/C Accuracy for °F is ±(0.05% of 80.00 mV + 1°F)

Source resolution is 1 °C or °F. Read resolution is 0.1 °C or °F

40