Altek 830-KP manual Read AC Voltages, Measure Thermocouple Sensors

Models: 830-KP

1 25
Download 25 pages 33.49 Kb
Page 13
Image 13

OPERATING INSTRUCTIONS

READ AC VOLTAGES

Choose this function to measure from 0.0 to 250.0 V True RMS.

1)Press the MILLIAMP/V push-button until mV, V, Vhi or VAC appear on the display

3)Press the DISPLAY/SOURCE/READ push-button until READ appears on the display

4)Press the TYPE/ENG UNITS pushbuuton to select VAC on the display

5)Connect the red READ (+) lead and the black READ (-) lead of the calibrator across the voltage to be measured.

Signals above or below those available for the currently selected range will be indicated by OVER and UNDER on the display.

CAUTION: Care should be used when measuring AC voltage. The included safety test probes or safety alligator clips should be used. Do not exceed voltage limits shown on calibrator.

OPERATING INSTRUCTIONS

MEASURE THERMOCOUPLE SENSORS

Choose this function to read a thermocouple. The input of the TechChek 830-KP is automatically cold junction compensated.

1)Disconnect the thermocouple from any instrument.

2)Press the T/C/RTD/pushbutton until the word TYPE & any T/C appear on the display.

3)Press the DISPLAY/SOURCE/READ push-button until READ and any T/C TYPE appear on the display

4)Use the proper thermocouple wire and corresponding miniature T/C connector to connect TechChek 830-KP to the thermocouple.

To Change the Thermocouple type

1)Press the TYPE/ENG UNITS push-button. The words SELECT T/C TYPE and the current selection appear on the display.

2)Press any UP/DOWN push-button until the desired T/C type appears.

3)Press the STORE/RESET push-button or wait 5 seconds to store the selection

AC Voltage Signal

Mains

Line Voltage

+ -

22

23

Page 13
Image 13
Altek 830-KP manual Read AC Voltages, Measure Thermocouple Sensors