OPERATING INSTRUCTIONS
SIMULATE 2-WIRE TRANSMITTERS
2-WIRE mA, 2-WIRE % (Percent of 4 to 20 mA) 2-WIRE DP %
Choose this function to simulate a
1.00to 24.00 milliamps. Operates in loops with power supply voltages from 3 to 45 VDC.
1)Disconnect existing
2)Press the MILLIAMP/V
3)Press the DISPLAY/SOURCE/READ
4)If SOURCE is on the display, press the TYPE/ENG UNITS push- button once to indicate
5)Press the mA/%/% DP FLOW
6)Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus
The simulated output of the
OPERATING INSTRUCTIONS
CALIBRATE VOLTAGE INPUTS
V, mV
Choose this function to provide an output from 0.00 mV to 110.00 mV and from 0.00 to 10.25 VDC. Current compliance up to 20 mA to provide the driving power to your voltage receivers.
1)Disconnect one or both input wires from the device to be calibrated
2)Press the MILLIAMP/V
3)Press the DISPLAY/SOURCE/READ
4)Press the TYPE/ENG UNITS
5)Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus
Output voltage is continuously adjustable with the UP/DOWN
To
Sensor
+ I N - REF +OUT-
Typical
Transmitter
(Disconnected) 10
Receiver
(Powers external
Power Supply
2 to 45 VDC
Voltage Receiver Input
Controller
Transmitter
Computer
Logger
DCS
11