OPERATING INSTRUCTIONS
COUNT FREQUENCIES
Choose this function to use the
To measure waveforms with amplitudes between 1 V and 10.25 V RMS use the low level inputs. Use the high voltage connection to read waveforms with amplitudes from 10.25 to 250.0 V RMS
1)Disconnect one or both input wires from the device to be calibrated
2)Press the FREQUENCY/PRESSURE pushbutton until any frequency signal (KHz, Hz or CPM) appears on the display.
3)Press the DISPLAY/SOURCE/READ
4)Press the TYPE/ENG UNITS
5)Connect the red READ lead of the calibrator to the plus (+) input of the device and the black READ lead to the minus
OPERATING INSTRUCTIONS
COUNT FREQUENCIES
Connection for signals with amplitudes below 10.25V RMS
Frequency Output Signal
Flowmeter
Flow Sensor
Variable Speed Drive
Controller
Transmitter
Connection for signals with amplitudes to 250 V RMS
Frequency Output Signal
Flow Meter
Line Frequency
Variable Speed Drives
28 | 29 |