Diagnostics and Service
3Use the temperature reading from Step 2 to determine the expected sensor resistance from Table 23.
4Using a calibrated ohmmeter, measure the actual resistance across the two sensor leads.
5Compare the expected resistance to the actual resistance.
6If the actual resistance value deviates substantially (more than 10%) from the expected resistance, replace the sensor.
Table 23: Temperature versus resistance
°F (°C) | Resistance in ohms | °F (°C) | Resistance in ohms |
613 | 113 (45) | 1195 | |
640 | 122 (50) | 1237 | |
668 | 131 (55) | 1279 | |
697 | 140 (60) | 1323 | |
727 | 149 (65) | 1368 | |
5 | 758 | 158 (70) | 1413 |
14 | 789 | 167 (75) | 1459 |
23 | 822 | 176 (80) | 1506 |
32 (0) | 855 | 185 (85) | 1554 |
41 (5) | 889 | 194 (90) | 1602 |
50 (10) | 924 | 203 (95) | 1652 |
59 (15) | 960 | 212 (100) | 1702 |
68 (20) | 997 | 221 (105) | 1753 |
77 (25) | 1035 | 230 (110) | 1804 |
86 (30) | 1074 | 239 (115) | 1856 |
95 (35) | 1113 | 248 (120) | 1908 |
104 (40) | 1153 |
|
|
Troubleshooting Humidity Sensors
The UVC is configured to use a
CAUTION
The humidity sensor is not protected against reversed polarity. Check carefully when connecting the device or damage can result.
Use the following procedure to troubleshoot a suspect sensor:
1Disconnect the sensors output voltage lead from the UVC analog input.
2Using some other calibrated humidity sensing device, take a humidity reading at the sensor location.
3Use the humidity reading from Step 2 determine the expected sensor voltage from Table 24.
4Using a calibrated
Wire color definitions: White = ground Yellow = output VDC Blue = supply VDC
5Compare the expected voltage to the actual voltage.
6If the actual voltage value deviates substantially (more than 10%) from the expected voltage, replace the sensor.
McQuay OM 751 | 45 |