Alarm Log

From the System Screen press the key adjacent to the AGM300 unit that you wish to view.

AGM Setup #1 Screen

Press the ALARM EVENT LOG key to display Alarm Log Screen.

Alarm Log Screen

Use the LEFT/RIGHT and UP/DOWN cursor keys to move through the log.

The Alarm log shows the last 20 alarm events. An alarm event is the occurrence of any alarm, any change in alarm level, or the clearing of any alarm. The alarm level is indicated by an L (leak), S (spill), or E (evacuate). The time and date of the alarm are displayed when the cursor is placed over the alarm.