Service Mode
When activated, the Service Mode will disable a specific AGM300 unit for a specified length of time. The default is 60 minutes. This time interval can be changed per Section Service Timeout on Page 40.
IMPORTANT: Note that while in the Service Mode no measurements are made, nor are any alarms activated.
From the System Screen (Page 31), select the key adjacent to the AGM unit you wish to place into its Service Mode. From AGM Setup Screen #1 (Page 37) press the SERVICE MODE ENTRY key twice within
3 seconds.
Note that the System Screen will show what monitors are in the Service Mode.
To exit the Service Mode, again press the SERVICE MODE ENTRY key twice within 3 seconds.
AGM Setup Screen #1
Press this key twice within
3 seconds to place the selected AGM unit into its Service Mode
System Screen (Service Mode)
Monitor in
Service
Mode
Instruction | 61 |