Bacharach AGM300, ADM800 manual System Screen Service Mode

Models: AGM300 ADM800

1 94
Download 94 pages 38.55 Kb
Page 69
Image 69

Service Mode

When activated, the Service Mode will disable a specific AGM300 unit for a specified length of time. The default is 60 minutes. This time interval can be changed per Section Service Timeout on Page 40.

IMPORTANT: Note that while in the Service Mode no measurements are made, nor are any alarms activated.

From the System Screen (Page 31), select the key adjacent to the AGM unit you wish to place into its Service Mode. From AGM Setup Screen #1 (Page 37) press the SERVICE MODE ENTRY key twice within

3 seconds.

Note that the System Screen will show what monitors are in the Service Mode.

To exit the Service Mode, again press the SERVICE MODE ENTRY key twice within 3 seconds.

AGM Setup Screen #1

Press this key twice within

3 seconds to place the selected AGM unit into its Service Mode

System Screen (Service Mode)

Monitor in

Service

Mode

Instruction 3015-4275

61

Page 69
Image 69
Bacharach AGM300, ADM800 manual System Screen Service Mode