Sensor Data Register

Register 0x0016h

R

82 bytes

 

 

 

 

 

 

Variable

Type

Length

Description

 

 

 

 

 

 

Pressure

Float

4

Manifold Pressure is PSIA

 

 

 

 

P_Volts

Float

4

Pressure sensor output Voltage

 

 

 

 

Vacuum_P

Float

4

Pressure with all value closed and pump on in PSIA

 

 

 

 

Ambient_P

Float

4

Absolute Ambient Pressure in PSIA

 

 

 

 

Box_T

Float

4

Enclosure Temperature in Degrees C

 

 

 

 

Box_T_Volts

Float

4

Box temp sensor output voltage

 

 

 

 

Bench_T

Float

4

Optical bench temperature in Degrees C

 

 

 

 

Bench_T_Volts

Float

4

Bench temp sensor output voltage

 

 

 

 

Bench_Z_T

Float

4

Optical bench temp in degrees C at last zero interval

 

 

 

 

Box_Z_T

Float

4

Box temp in degrees C at last zeroing

 

 

 

 

PkPk_int

UI

2

Current peak to peak A/D counts from detector

 

 

 

 

PkPk

Float

4

Current peak to peak voltage from detector

 

 

 

 

Ave_PkPk

Float

4

Running average voltage from detector

 

 

 

 

Zero_PkPk

Float

4

Voltage measured at last zeroing

 

 

 

 

Noise

Float

4

Largest Change in running average

 

 

 

 

AU

Float

4

Current absorbance value

 

 

 

 

Ave_AU

Float

4

Running Average of absorbance level

 

 

 

 

Bench_PPM

Float

4

PPM in bench based on zone gas selection (uncorrected

 

 

 

for pressure and temperature)

 

 

 

 

STP_PPM

Float

4

PPM corrected to STP (1 atm, 25 deg C)

 

 

 

 

Bench_UML

Float

4

Micromoles/liter in bench (uncorrected0

 

 

 

 

Ambient_UML

Float

4

Micromoles/liter corrected to ambient pressure

 

 

 

 

 

 

76

Instruction 3015-4275

Page 84
Image 84
Bacharach ADM800, AGM300 manual Stpppm

AGM300, ADM800 specifications

The Bacharach ADM 800 and AGM 300 are advanced analyzers designed for efficient monitoring and detection of gases in various applications. Known for their reliability and accuracy, these instruments cater to commercial and industrial sectors, ensuring safety and compliance with environmental regulations.

The Bacharach ADM 800 is a portable gas detection device renowned for its ability to detect multiple refrigerants. With a user-friendly interface, it features a high-resolution display that provides real-time data on refrigerant concentrations, allowing technicians to quickly identify leaks during maintenance operations. One of its standout technologies is a highly sensitive thermal conductivity sensor, which enables the ADM 800 to differentiate between various hydrocarbon refrigerants with precision.

Equipped with data logging capabilities, the ADM 800 can store and retrieve measurement data, making it an invaluable tool for tracking refrigerant trends over time. The device supports USB connectivity for easy transfer of data to computers, facilitating detailed reporting and analysis. Its robust design incorporates a protective casing that ensures durability in challenging work environments, making it ideal for both field use and stationary applications.

On the other hand, the Bacharach AGM 300 is specifically crafted for continuous gas monitoring in HVAC systems. This stationary device excels in measuring concentrations of gases such as CO, CO2, and various refrigerants, ensuring optimal performance and compliance. The AGM 300 uses advanced electrochemical sensors for precise gas measurements, providing real-time feedback and alerting users to any hazardous conditions.

One of the notable features of the AGM 300 is its ability to integrate seamlessly with building management systems, enabling centralized control and monitoring of gas levels across multiple locations. This connectivity enhances the safety of industrial and commercial spaces by allowing for automated responses to detected gas levels, which can significantly mitigate risks associated with gas leaks.

Both models emphasize user-centric design, with intuitive interfaces and clear displays, making them accessible even to those with minimal training. Additionally, their low maintenance requirements and long sensor life contribute to reduced operational costs.

In conclusion, the Bacharach ADM 800 and AGM 300 are exemplary solutions for gas detection and monitoring. Their advanced technologies, robust features, and user-friendly designs make them indispensable for ensuring safety and regulatory compliance in various environments.