ABOUT DISPLAY FUNCTIONS

Native, uncorrected panels will display all grayscale/color levels at an equal luminance. Studies have shown however, that in medical images certain grayscale/color parts contain more diagnostic information then others. To respond to these conclusions, display functions have been defined. These functions emphasize on these parts containing crucial diagnostic information by correcting the native panel behavior.

Native

If you select Native, the native panel

 

behavior will not be corrected.

 

 

Dynamic Gamma 1.8

These are gamma functions that are shifted

Dynamic Gamma 2.2

to take into account the non-zero

 

luminance of an LCD panel when driven

 

with a “black” signal.

 

They are especially useful in CT applications

 

to improve the perception of low

 

Haunsfield values.

 

 

DICOM

DICOM (Digital Imaging and

 

Communications in Medicine) is an

 

international standard that was developed

 

to improve the quality and communication

 

of digital images in radiology.

 

In short, the DICOM display function results

 

in more visible grayscales in the images.

 

Barco recommends selecting the DICOM

 

display function for most medical viewing

 

applications.

 

 

User

This display function will be automatically

 

selected when display functions are

 

defined by MediCal QAWeb.

 

 

Gamma 1.8

Select one of these display functions in

Gamma 2.2

case the display is to replace a CRT display

 

with a gamma of 1.8 or 2.2 respectively.

 

 

32 Coronis Fusion 10MP