Reading rooms

ABOUT READING ROOMS

Reading rooms can only be selected when the DICOM display function is selected. Therefore, please refer to “Display functions” on page 32 to correctly set the display function.

The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms. Each of these reading rooms are defined by following parameters:

the maximum light allowed in this type of room

the preset ambient light correction value for this reading room

These parameters are stored in your display and determine the preset ambient light correction value to take into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled. Please refer to “Ambient Light Compensation (ALC)” on page 33 to enable ALC.

The available reading rooms for your CORONIS FUSION 10MP display are:

CR/DR/MAMMO

Corresponds to light conditions in

 

diagnostic reading rooms for computed

 

radiology, digital radiology or

 

mammography. This setting has the lowest

 

maximum ambient light.

 

 

CT/MR/NM

Corresponds to light conditions in

 

diagnostic reading rooms for computed

 

tomography, magnetic resonance or

 

nuclear medicine scans.

 

 

Staff Office

Corresponds to light conditions in office

 

rooms.

 

 

Clinical Viewing

Corresponds to light conditions in

Room

diagnostic reading rooms for clinical

 

viewing.

 

 