MCA Mode Command:
Command name | CMD_SET_MODE |
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Description | Depending on the value of the first parameter (mod): |
|
|
|
|
|
| ||||||||
| MODE_MCA | = 0 |
|
|
|
|
|
|
|
|
|
| |||
| MODE_MCS | = 1 |
|
|
|
|
|
|
|
|
|
| |||
| the command set MMCA to the required mode by activating the previous specified setup parameter. | ||||||||||||||
| The command will be ignored and responded with a error messages, if a measurement is still |
| |||||||||||||
| running. |
|
|
|
|
|
|
|
|
|
|
|
|
| |
Command syntax |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Format | integer |
|
| integer |
| integer |
| long |
|
|
| integer |
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
Parameter | preamble |
| command | mod |
| 0 |
|
|
| end flag |
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Byte String (HEX) | A5 |
| 5A |
| 45 | 00 | mod | 00 | 00 | 00 | 00 | 00 | B9 |
| 9B |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Used setup value | coarse_gain, fine_gain, mca_input_pol, mca_input_adc |
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Affected setup value | mca_mode, stab_state |
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
MCA state | STATE_READY |
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MCA Clear Command: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
Command name | CMD_CLEAR |
|
|
|
|
|
|
|
|
|
|
| |||
|
| ||||||||||||||
Description | A running measurement is stopped immediately. Depending on the value of the first parameter | ||||||||||||||
| (clear): |
|
|
|
|
|
|
|
|
|
|
|
|
| |
| CLEAR_MEMORY = 0 |
|
|
|
|
|
|
|
|
|
| ||||
| CLEAR_TIME | = 1 |
|
|
|
|
|
|
|
|
|
| |||
| CLEAR_ROI | = 2 |
|
|
|
|
|
|
|
|
|
| |||
| CLEAR_ALL | = 3 |
|
|
|
|
|
|
|
|
|
| |||
| the command clears the spectrum, the time information, the ROI limits exclusively or all together. | ||||||||||||||
Command syntax |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Format | integer |
|
| integer |
| integer |
| long |
|
|
| integer |
| ||
|
|
|
|
|
|
|
|
|
|
|
| ||||
Parameter | preamble |
| command | clear |
| 0 |
|
|
| end flag |
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Byte String (HEX) | A5 |
| 5A |
| 44 | 00 | clear | 00 | 00 | 00 | 00 | 00 | B9 |
| 9B |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Used setup value | None |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| ||||||
Affected setup value | real time, dead time, ROI begin, ROI end |
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||
MCA state | STATE_READY |
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
59