Braun MCA166-USB user manual

Models: MCA166-USB

1 72
Download 72 pages 41.23 Kb
Page 24
Image 24

MCA166-USB

12.Bad resolution in a HPGe spectrum or from a test generator, specially at high amplification factors

Check electrical noise and EMC compatibility of surroundings. Some laptops and also some switch mode power supply battery chargers may disturb. Remove all connections from the MCA to ground or to mains voltages. Try a different computer / laptop / palmtop.

Winding the cable to the detector / computer several times around a ferrite ring also may help

13.While opening a spectrum, the MCA program tells "Wrong data format" or "data format error"

spectrum was created by another program or another program version. Check results, just ignore.

14.too high count rate

Keep a bigger distance to radiation source. Try a less sensitive detector. Watch for your personal safety!

Check detector

24

Page 24
Image 24
Braun MCA166-USB user manual