Braun MCA166-USB user manual Cmdqueryuserdata, Cmdsystemdata, Cmdquerycentroid

Models: MCA166-USB

1 72
Download 72 pages 41.23 Kb
Page 70
Image 70

MCA166-USB

Command name

CMD_QUERY_USER_DATA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command reads a 32 bit value from the MCA parameter memory.

 

 

 

 

 

e: 0 ... 63 parameter number

 

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

integer

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

command

e

 

0

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

5E

00

e

00

00

00

00

00

B9

 

9B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Used setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Result

see MMCA_QUERY_USER_DATA function in DLL description.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_SYSTEM_DATA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command reads the special MCA data.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

integer

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

command

00

 

End

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

62

00

00

00

00

00

00

00

B9

 

9B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Used setup value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Result

see MMCA_QUERY_SYSTEM_DATA function in DLL description.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_QUERY_CENTROID

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command reads the centroid of a ROI (beg<end, LLD<=beg, end<=ULD, end-beg<250).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

beg

 

End

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

5F

00

begl

begh

endl

endh

00

00

B9

 

9B

Used setup value

channels, lld, uld

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Result

see MMCA_QUERY_CENTROID function in DLL description.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 