MCA166-USB
Name | MMCA_QUERY_UF6_INFO |
|
|
|
|
Description | The command reads current dead time, real time the integral, begin and end of the UF6 ROIs | |
|
|
|
Parameter | - |
|
|
|
|
Result array | Real time | long value |
|
|
|
Byte Offset 4 | Dead time [msec] | long value |
|
|
|
Byte Offset 8 | Integral ROI 1 | long value |
|
|
|
Byte Offset 12 | Integral ROI 2 | long value |
|
|
|
Byte Offset 16 | Integral ROI 3 | long value |
|
|
|
Byte Offset 20 | Begin ROI 1 | long value |
|
|
|
... | ... | long value |
|
|
|
Byte Offset 40 | End ROI 3 | long value |
|
|
|
Return Code | see MMCA_RESET command |
|
|
|
|
Name | MMCA_QUERY_UF6_ROIS |
|
|
|
|
Description | The command reads the begin and end of the 3 ROIs | |
|
|
|
Parameter | - |
|
|
|
|
Result array | Begin ROI 1 | long value |
|
|
|
Byte Offset 4 | End ROI 1 | long value |
|
|
|
Byte Offset 8 | Begin ROI 2 | long value |
|
|
|
Byte Offset 12 | End ROI 2 | long value |
|
|
|
Byte Offset 16 | Begin ROI 3 | long value |
|
|
|
Byte Offset 20 | End ROI 3 | long value |
|
|
|
Return Code | see MMCA_RESET command |
|
|
|
|
Name | MMCA_QUERY_CENTROID |
| |
|
|
|
|
Description | The command reads peak centroid of the specified ROI | ||
|
|
|
|
1. Parameter | Peak ROI begin | integer value | LLD<= begin < end |
|
|
|
|
2. Parameter | Peak ROI end | integer value | begin < end <= ULD, |
|
|
|
|
Result | Peak centroid | float value |
|
|
|
|
|
Return Code | see MMCA_RESET command |
| |
|
|
|
|
Name | MMCA_QUERY_ON_LINE |
|
|
Description | The command checks the communication between the MCA and the PC. |
|
|
Parameter | - |
|
|
Results | - |
|
|
Return Code | see MMCA_RESET command |
|
|
56