Braun MCA166-USB user manual Cmdsettdf, CMDSETUF6ROI, Cmdsetuserdata

Models: MCA166-USB

1 72
Download 72 pages 41.23 Kb
Page 68
Image 68

MCA166-USB

MCA Data Setup Commands:

Command name

CMD_SET_TDF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command sets the TDF parameter.

 

 

 

 

 

 

 

 

 

tdf:

TDF(100 ... 3000, default 800)

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

tdf

 

0

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

61

00

tdfl

tdfh

0

0

0

0

B9

 

9B

Used setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

tdf

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_SET_UF6_ROI

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command sets the begin and the end of a ROI used by the other UF6 commands.

 

 

r:

 

ROI number

 

 

 

 

 

 

 

 

 

 

 

b:

 

LLD<= b < e

 

 

 

 

 

 

 

 

 

 

 

e:

 

b < e <= ULD, (e-b) < 500

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

 

integer

 

integer

 

integer

 

integer

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

Preamble

 

 

command

r

 

b

 

e

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

64

00

r

00

bl

bh

el

eh

B9

 

9B

Used setup value

LLD, ULD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

roi_begin[n], roi_end[n]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_SET_USER_DATA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command stores a 32 bit value(f) in the MCA parameter memory(64 entries).

 

 

 

 

e: 0 ... 63

parameter number

 

 

 

 

 

 

 

 

 

p: 0 ...

 

 

long or float parameter

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

integer

 

long or float

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

e

 

P

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

57

00

e

 

00

pl

pl

ph

ph

B9

 

9B

Used setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 