MCA166-USB
Name | MMCA_QUERY_POWER |
|
|
|
|
|
|
|
|
|
|
Description | The command reads the MMCA power state. |
|
|
|
|
|
|
|
|
|
|
Parameter | - |
|
|
|
|
|
|
|
|
|
|
Result array | Battery current [mA] | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 4 | HV primary current [mA] | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 8 | +12 primary current [mA] | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 12 | long value |
|
|
| |
|
|
|
|
|
|
Byte Offset 16 | +24 primary current [mA] | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 20 | long value |
|
|
| |
|
|
|
|
|
|
Byte Offset 24 | Battery voltage [mV] | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 28 | HV [V]27 | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 32 | HV state | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 36 |
|
|
|
|
|
|
|
|
|
|
|
Byte Offset 40 | Current high voltage [V] |
|
|
|
|
|
|
|
|
|
|
Byte Offset 44 |
|
|
|
|
|
|
|
|
|
|
|
Byte Offset 48 | Power Switches | long value | 0X80 | ON | |
|
|
| 0X40 | +24V | ON |
|
|
| 0X20 | ON | |
|
|
| 0X10 | +12V | ON |
|
|
| 0X02 | HV | ON |
Byte Offset 52 | Charger current [mA] | long value |
|
|
|
|
|
|
|
|
|
... |
|
|
|
|
|
|
|
|
|
|
|
Byte Offset 64 | Battery current [mA] at stop | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 68 | HV primary current [mA] at stop | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 72 | +12 primary current [mA] at stop | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 76 | long value |
|
|
| |
|
|
|
|
|
|
Byte Offset 80 | +24 primary current [mA] at stop | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 84 | long value |
|
|
| |
|
|
|
|
|
|
Byte Offset 88 | Battery voltage [mV] at stop | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 92 | HV [V]28 at stop | long value |
|
|
|
|
|
|
|
|
|
Byte Offset 116 | Charger current [mA] at stop | long value |
|
|
|
|
|
|
|
|
|
... |
|
|
|
|
|
|
|
|
|
|
|
Byte Offset 128 | mca_state29 | integer value |
|
|
|
|
|
|
|
|
|
Return Code | see MMCA_RESET command |
|
|
|
|
|
|
|
|
|
|
27multiply value with 1.2 V
28multiply value with 1.2 V
29see MMCA_QUERY_SYSTEM_DATA
50