MCA166-USB

MCA Measurement Setup Commands:

Command name

CMD_SET_ADC_RES_DISCR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command set the ADC resolution(res: 128, 256 … 4096), the LLD(Low level discriminator) and

 

the ULD(Upper level discriminator) to the values of its three parameter.

 

 

 

 

 

The command will be ignored and responded with a error messages, if one ore more parameter are

 

invalid or a measurement is still running.

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

integer

 

long

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

res

 

LLD, ULD

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

46

00

resl

resh

LLDl

LLDh

ULDl

ULDh

B9

 

9B

Used setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

channels, lld, uld, stab_state

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_SET_PRESETS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command sets the choice(pre)

 

 

 

 

 

 

 

 

 

PRESET_NONE

= 0

 

 

 

 

 

 

 

 

 

 

 

PRESET_REAL

= 1

 

 

 

 

 

 

 

 

 

 

 

PRESET_LIVE

= 2

 

 

 

 

 

 

 

 

 

 

 

PRESET_INT

= 3

 

 

 

 

 

 

 

 

 

 

 

PRESET_AREA

= 4

 

 

 

 

 

 

 

 

 

 

 

and value(val) of the preset. The parameter val must be lower then 65536 for PRESET_REAL and

 

PRESET_LIVE! If a measurement is running, only val can be changed.

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

pre

 

val

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

48

00

prel

preh

vall

...

...

valh

B9

 

9B

Used setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

presets, preset_value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_SET_ROI

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command sets the begin and end of the ROI for the preset integral and area.(LLD <= begin <

 

end and LLD < end <= ULD)

 

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

beg

 

end

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

49

00

begl

begh

endl

endh

00

00

B9

 

9B

Used setup value

LLD, ULD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

roi_begin, roi_end

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

60

Page 60
Image 60
Braun MCA166-USB Cmdsetadcresdiscr, Lld, Uld, Cmdsetpresets, Presetnone Presetreal Presetlive Presetint Presetarea

MCA166-USB specifications

The Braun MCA166-USB is a compact and versatile charging device designed for modern lifestyles. This product expertly combines functionality with portability, making it an essential accessory for people on the go. One of the key features of the Braun MCA166-USB is its dual charging capability. It comes equipped with multiple USB ports, allowing users to charge various devices simultaneously, from smartphones to tablets. This multi-device functionality eliminates the need for multiple chargers cluttering your workspace or travel bag.

The device supports fast charging technology, which is particularly beneficial for users who need quick power boosts. With a maximum output of 2.4A per port, the Braun MCA166-USB ensures that your devices are energized swiftly, reducing downtime and keeping you connected during busy days. Additionally, the device is compatible with a wide range of charging protocols, offering versatility for different brands and models.

The Braun MCA166-USB features an intelligent charging system that recognizes the connected device and delivers the optimal charging current. This not only enhances the efficiency of charging but also protects your devices from any potential damage caused by overcharging.

Constructed with a sleek and sturdy design, the Braun MCA166-USB is not just about functionality; it is also aesthetically pleasing. Its compact size makes it perfect for portability, easily fitting into a pocket or small bag. The durability of the materials used ensures that the charger withstands the rigors of travel while remaining stylish.

Another highlight of the Braun MCA166-USB is its safety features. It incorporates multiple protection mechanisms, including short-circuit prevention and temperature control, ensuring a safe charging environment for your devices.

In summary, the Braun MCA166-USB emerges as an innovative solution for those seeking a reliable charging option. Its robust features, such as dual USB ports, fast charging capabilities, intelligent charging control, and safety measures, make it an ideal choice for anyone looking to stay powered up while balancing a busy lifestyle. With its sleek design and durable construction, the Braun MCA166-USB undoubtedly stands out as a multifaceted charging device for the modern user.