Braun MCA166-USB user manual Cmdsetstabparam

Models: MCA166-USB

1 72
Download 72 pages 41.23 Kb
Page 66
Image 66

MCA166-USB

MCA Stabilisation Commands:

Command name

CMD_SET_STABILISATION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command sets the peak stabilisation.

 

 

 

 

 

 

 

 

 

fl:

0

 

 

 

turns stabilisation off

 

 

 

 

 

 

 

 

 

1

 

 

 

turns stabilisation on

 

 

 

 

 

 

 

 

 

rb+3<fl<re-3

 

stabilisation to channel fl

 

 

 

 

 

 

 

rb:

Peak ROI begin (LLD<= begin < end)

 

 

 

 

 

 

 

 

re:

Peak ROI end (begin < end <= ULD, (end-begin) < 250)

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

 

fl

 

rb

 

re

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

4D

00

 

fll

flh

rbl

rbh

rel

reh

B9

 

9B

Used setup value

channels, lld, uld

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

stab_state, stab_roi_begin, stab_roi_end

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_SET_STAB_PARAM

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command sets the stabilisation parameters.

 

 

 

 

 

 

 

 

st:

time interval (default 10)

 

 

 

 

 

 

 

 

 

 

sa:

area (default 25000)

 

 

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

Preamble

 

command

 

st

 

sa

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

67

00

 

stl

sth

sal

sal

sal

sah

B9

 

9B

Used setup value

-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

stab_time, stab_area

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

66

Page 66
Image 66