Braun MCA166-USB user manual Cmdsetpzctimeoffset

Models: MCA166-USB

1 72
Download 72 pages 41.23 Kb
Page 65
Image 65

Command name

CMD_SET_PZC

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command performs the Pole Zero Cancellation.

 

 

 

 

 

 

 

 

pv: 0 ... 2499

PZC control voltage

 

 

 

 

 

 

 

 

 

 

pm = 0

 

 

set PCZ only

 

 

 

 

 

 

 

 

 

 

 

pm > 0

 

 

set PZC and measure PZC offset

 

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

integer

 

long

 

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

pm

 

Pv

 

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

58

00

pml

pmh

pvl

pvh

00

 

00

B9

 

9B

Used setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

pzc_value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Results

If pm > 0 the function returns with the number of counts and the - PZC offset.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command name

CMD_SET_PZC_TIME_OFFSET

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

The command sets the time offset for Pole Zero Cancellation.

 

 

 

 

 

 

t1: 0 ... 31

Time offset for low shaping time

 

 

 

 

 

 

 

t2: 0 ... 31

Time offset for high shaping time

 

 

 

 

 

 

Command syntax

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format

integer

 

 

integer

 

 

integer

 

long

 

 

 

integer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

preamble

 

command

 

t1

 

t2

 

 

 

end flag

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Byte String (HEX)

A5

5A

 

60

00

 

t1

00

t2

00

00

00

B9

 

9B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Used setup value

None

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Affected setup value

pzc_dtc_1_offset, pzc_dtc_3_offset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MCA state

STATE_READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 