Section 5. Programming the Datalogger
x – this filter marks the start of the data set. In this case the data set is two floating point numbers.
f – this filter searches for the first ASCII floating point number to convert.
f – this filter searches for the second ASCII floating point number to convert.
*Table 1 Program
01: 2 Execution Interval (seconds)
;do filter set-up  only if flag one is clear.
1:If Flag/Port (P91)
1:21 Do if Flag 1 is Low
2: 30  | Then Do  | 
;set up string filter.
2:
1: 1  | Reps  | 
2: 0  | Address  | 
3: 1  | Send/Receive Port 1  | 
4: 2054  | Command  | 
5: 9200  | 1st Parameters  | 
6: 0  | 2nd Parameters  | 
7: 0  | Values per Rep  | 
8:0 Loc [ _________ ]
9:1.0 Mult
10:0.0 Offset
;delay required by filter set-up  command.
3:Excitation with Delay (P22)
1:1 Ex Channel
2:0 Delay W/Ex (units = 0.01 sec)
3:1 Delay After Ex (units = 0.01 sec)
4:0 mV Excitation
;set flag one high so the filter set-up  is only done once.
4:Do (P86)
1:11 Set Flag 1 High
;end of IF.
5:End (P95)
;poll SDM-SIO4  to see if data is available.
6:
1:  | 1  | Reps  | 
  | 
2:  | 0  | Address  | 
  | 
3:  | 1  | Send/Receive Port 1  | |
4:  | 1  | Command  | 
  | 
5:  | 0  | 1st Parameters  | 
  | 
6:  | 0  | 2nd Parameters  | 
  | 
7:  | 1  | Values per Rep  | 
  | 
8:  | 2  | Loc [ poll  | ]  | 
9:  | 1.0  | Mult  | 
  | 
10:  | 0.0  | Offset  | 
  |