Immediate
0 to time remaining in the infusion period, in 10 minute increments (defaults to the currently programmed
AutoLock
Not In Use, LL1 or LL2
Time
00:00 to 23:59
Air Detector
Turned On or Turned Off
Event Log
0 to 500 events
Extended History
Up to 48 hours in 1 hour increments
Biomed Toolbox SpecificationsMicrograms
On or Off
Extended History
On or Off
Max Dose per Hour On or Off
(PCA software rev. 6210G or higher only)
PM (Preventive Maintenance) Reminder
1 to 24 months in 1 month increments,
Not In Use
Custom Lock Level Code
Date Format
US Standard (mm/dd/yy) or European
Standard (dd/mm/yy)
Power Source Display
Always display or only Low Battery
Upstream Sensor On or Off (Model 6101 only)
Air Detector Required
Required or Not Required
Compatible Reservoirs andAdministration Sets•
•CADD® administration set with integral
(allows use of flexible plastic bag or sterile vial with injector)
•CADD® administration set with add on anti- siphon valve and bag spike (allows for gravity priming before attaching the add on anti- siphon valve)
•
Deltec provides a Remote Dose Cord for the PCA delivery mode, which is an extension of the DOSE key. The push button switch is a Single Pole Double Throw (SPDT) which operates in the same manner as the DOSE key. When the Remote Dose Cord is attached to the pump, the patient may press either the Remote Dose button or the DOSE key to receive a Demand Dose. The clinician may also use either the Remote Dose button or the DOSE key to deliver a clinician bolus. For easy access, the Remote Dose cord may be fastened to the patient’s clothing or bedsheet with the attached clip.
NOTE:To detach the Remote Dose cord from the pump, grasp the Remote Dose cord connector and pull back using a straight, steady motion. Do not twist or turn the connector, or use any instrument to remove it.
For additional specifications refer to the Operator’s Manual provided with the product.