Options Specifications

Immediate Taper-Down (TPN)

0 to time remaining in the infusion period, in 10 minute increments (defaults to the currently programmed Taper-Down)

AutoLock

Not In Use, LL1 or LL2

Time

00:00 to 23:59

Air Detector

Turned On or Turned Off

Event Log

0 to 500 events

Extended History

Up to 48 hours in 1 hour increments

Biomed Toolbox Specifications

Micrograms

On or Off

Extended History

On or Off

Max Dose per Hour On or Off

(PCA software rev. 6210G or higher only)

PM (Preventive Maintenance) Reminder

1 to 24 months in 1 month increments,

Not In Use

Custom Lock Level Code

Date Format

US Standard (mm/dd/yy) or European

Standard (dd/mm/yy)

Power Source Display

Always display or only Low Battery

Upstream Sensor On or Off (Model 6101 only)

Air Detector Required

Required or Not Required

Compatible Reservoirs andAdministration Sets

50-ml or 100-ml Medication Cassette reser- voir, used with the CADD® extension set with anti-siphon valve.

CADD® administration set with integral anti-siphon valve, with or without bag spike

(allows use of flexible plastic bag or sterile vial with injector)

CADD® administration set with add on anti- siphon valve and bag spike (allows for gravity priming before attaching the add on anti- siphon valve)

CADD-Prizm®high volume administration set with add on anti-siphon valve, with or without air-eliminating filter

Remote Dose Cord

Deltec provides a Remote Dose Cord for the PCA delivery mode, which is an extension of the DOSE key. The push button switch is a Single Pole Double Throw (SPDT) which operates in the same manner as the DOSE key. When the Remote Dose Cord is attached to the pump, the patient may press either the Remote Dose button or the DOSE key to receive a Demand Dose. The clinician may also use either the Remote Dose button or the DOSE key to deliver a clinician bolus. For easy access, the Remote Dose cord may be fastened to the patient’s clothing or bedsheet with the attached clip.

NOTE:

To detach the Remote Dose cord from the pump, grasp the Remote Dose cord connector and pull back using a straight, steady motion. Do not twist or turn the connector, or use any instrument to remove it.

For additional specifications refer to the Operator’s Manual provided with the product.