Given

0 to 99999.99 in 0.01 unit increments

Clinician Bolus

0.1ml to 20.00 ml (or mg or mcg equiva- lent) (See Tables 2, 3 and 4 for scroll ranges) Delivery rate (Continuous Rate + Clinician Bolus): 125 ml/hr nominal

Continuous Delivery Mode Specifications

Reservoir Volume

1 to 9999 or Not In Use; programmable in

1 ml increments, displayed in 0.1 ml increments

Default: 1 ml

Continuous Rate

0.1to 350 ml/hr in the following increments:

0.1for values between 0.1 and 100 1 for values between 100 and 350 Default: 0 ml/hr

Use 9-volt battery for rates up to 250 ml/ hr; use power pack or AC adapter for rates up to 350 ml/hr

Given

0 to 99999.9 in 0.1 unit increments

TPN Delivery Mode Specifications

Reservoir Volume

10.0to 9990 or Not In Use; program- mable in 1 ml increments, displayed in 0.1 ml increments

Default: 10 ml

Infusion Volume

10 to 9990 ml in 10 ml increments: Default: 10 ml

Infusion Period

0 hrs 10 min to 99 hrs 50 min in 10 minute increments:

Default: 1 hr 0 min

Taper-Up Period

0 hrs 0 min to 99 hrs 40 min in 10 minute increments:

Default: 0 hrs 0 min

Taper-Down Period

0 hrs 0 min to 99 hrs 40 min in 10 minute increments:

Default: 0 hrs 0 min

Plateau Rate

Calculated by pump; 10 to 350 ml/hr

Use 9-volt battery for rates up to 250 ml/hr; use power pack or AC adapter for rates up to 350 ml/hr

KVO Rate

Calculated by pump: 1/10 of Plateau Rate up to 5 ml/hr

Given

0 to 99999 in 0.1 unit increments

Intermittent Delivery Mode Specifications

Reservoir Volume

1 to 9999 or Not In Use; programmable in

1 ml increments, displayed in 0.1 ml increments Default: 1 ml

Dose Volume

0.1to 1000 ml in the following increments:

0.1for values between 0.0 and 100 1 for values between 100 and 1000 Default: 0.0 ml

Dose Duration

1 min to 24 hrs in the following increments: 1 minute for values between 1 min and 10 min 5 minutes for values above 10 min Default: 30 min

Duration is limited by Dose Volume so that rate does not exceed 350 ml/hr. Use 9-volt battery for rates up to 250 ml/hr; use power pack or AC adapter for rates up to 350 ml/hr

Dose Cycle

10 min to 96 hrs in 5 minute increments: Default: 4 hrs

KVO Rate

0 to 10 ml/hr in 0.1 ml/hr increments Default: 0 ml/hr

Next Dose Start Time

10 min to 96 hrs X min (where X equals a 10 minute increment in the 96th hour) or Imme- diate; programmable in 10 minute increments Default: Immediate

Given

0 to 99999.9 in 0.1 unit increments

Time Remaining

Dose and Cycle display in 1 minute increments

6