delivery volume. This is the inaccuracy volume.

8.Divide the inaccuracy volume by the intended delivery volume and multiply by 100. This is the accuracy error percentage. (See Table 15.)

9.If the accuracy error percentage is greater than ±6%, repeat the test with a new reservoir. If the pump fails a second time, call Deltec’s Customer Service Department.

Intended

Actual

Inaccuracy

Accuracy

Accuracy

Delivery

Delivery

Volume

Error

Error Percentage

Volume

Volume

 

 

 

 

 

 

 

 

20 ml

19.5 ml

-0.5 ml

-0.5/20.0 ml =-0.025

-0.025 x 100 = ±2.5%

 

 

 

 

 

Table 15. Volumetric percentage calculation