CADD-Prizm® Pump Cleaning and Functional Testing Checklist
The following checklist is provided as a guide only to assist in establishing documentation of cleaning and functional testing for the
Serial # _________________ Reference Number _________________________ Date _________________________
(Refer to the Technical Manual procedures.)
I. | Cleaning Completed | Yes | No |
|
II. | Visual Inspection |
|
|
|
| LCD | Lock |
| Data In/Out Jack |
| Occlusion Sensor Seals | Cassette Sensors (3) | Air Detector or Port Cover | |
| Valves and Expulsors | Keyboard | Pump Housing | |
| Reservoir Hinge Area | Indicator Light | Battery Door | |
| Latch | Power Jack | Battery Compartment | |
III. Mechanical Inspection |
|
|
| |
| Keyboard | Cassette Latch |
| |
| Battery Door | Cassette Lock |
| |
IV. Functional Inspection |
|
|
| |
| | RES VOL Alarm | Dose Given / Attempted | |
| Latch / Lock | Stop / Start | MG Given | |
| Cassette Sensor | Activation Timing | Air Detector | |
| LCD | Dose Key | Printer Tests | |
| Motor / Gear Train | Remote Dose Cord |
|
V.Occlusion Tests (Only need to perform 1 or 2; not both) Downstream Option 1: Activations Before Alarm ___________
Downstream Option 2: High Pressure Alarm At ____________ psi
Upstream Occlusion Sensor Test
Pass ________ Fail ___________
VI. Accuracy Testing (Only need to perform 1 or 2; not both)
Volumetric Accuaracy Test
| Intended |
|
| Actual |
|
| Inaccuracy | Accuracy |
|
| Accuracy | |||
| Delivery |
|
| Delivery |
|
| Volume | Error |
|
|
| Error | ||
| Volume |
|
| Volume |
|
|
|
|
|
|
| Percentage | ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ml |
|
|
| ml |
| ml |
|
|
|
| % | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Gravimetric Accuracy Test |
|
|
|
|
|
|
|
|
|
|
| ||
|
| Amount | Intended | Inaccuracy |
| Accuracy |
| Accuracy | ||||||
| Weights | Weight |
|
| Delivered | Delivery | Volume |
| Error |
| Error | |||
|
|
|
|
|
|
|
| Volume |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| g |
|
| g |
|
| ml | ml | ml |
|
|
|
| % |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
31