Mechanical Inspection

Press each key on the keyboard. Each key should have a distinctive dome feeling. The keys should not feel flat.

Attach the battery door. The battery door should fit snugly in place when it is closed on the pump.

Attach either a 50- or 100-ml Medication Cassette reservoir or a CADD® administra- tion set to the pump. Check for smooth operation and a definite “feel” when the latch pulls the reservoir or administration set firmly against the bottom of the pump. The mark on the latch should be aligned with the solid dot.

Lock the device by inserting a key into the lock and turn counterclockwise until the mark lines up with the solid dot.

Gently twist and pull on the cassette to make sure it is firmly attached.

NOTE:

In the PCA delivery mode, the cassette must be locked in order to start the pump.

22