detector. An infrared light source is mounted so that its light beam illuminates the infrared detector. An opaque flag is mounted concentrically to the camshaft and rotates with it between the infrared light source and detector. When the flag interrupts the light beam, the output of the detector is sensed by the microprocessor via an input port bit. Power to the infrared LED light source is controlled by the motor driver circuit and is off when the motor is not running to conserve battery life.
In the microprocessor software, multiple checks are made on motion of the camshaft. When the motor is commanded to start, the infrared sensor must show that half a revolution has occurred within five seconds and that the motor has stopped when half a rotation was completed. In addition, no camshaft rotation can take place when the motor has not been commanded to run.
Power Circuitry
Power for the pump is normally supplied by a
The motor driver circuit power is taken directly from the battery, but the microprocessor and its associated circuitry requires closely regulated and filtered
Voltage Reference Circuit
A voltage reference circuit provides a constant DC voltage to the microprocessor Analog to Digital Converter (ADC). By reading this input and comparing the value to a predetermined range, the microprocessor can validate the accuracy of the
Voltage | CADD® Pump Status |
Trip Point* |
|
|
|
>7.0V | No alarm |
Transition to low battery | |
| condition; battery low |
| message appears; 3 beeps |
| every 5 min.† |
Transition to depleted | |
| battery condition; battery |
| depleted message appears; |
| continuous alarm†† |
Hardware reset occurs. | |
| Pump continues to indicate |
| depleted battery condition. |
*Voltage ranges are due to component tolerances. Actual trip values are guaran- teed to be
Table 12.
†The pump emits 3 beeps every 5 minutes, and the message “9 Volt Battery Low” appears on the pump’s display, indicating that the battery power is low, but the pump is operable.
††The pump emits a continuous,
Pumping Mechanism
The pumping mechanism is linear peristaltic with two active valves. Pumping occurs when the expulsor presses on the reservoir pump tubing in sequence with the inlet and outlet valves. At rest, the outlet valve is pressing down fully on the tubing and the expulsor and inlet valve are retracted. (See Figure 7.)
14