Intermittent Delivery Battery Life

Note: Results are without air detector.

Volume

DurationCycleKVO

Life

Volume

23.5

1:00 hr

5:00 hr

0.2 ml/hr

117 hrs

477 ml

61

1:00 hr

6:00 hr

0.2 ml/hr

71 hrs

733 ml

125

1:00 hr

6:00 hr

0.2 ml/hr

38 hrs

798 ml

200

1:00 hr

12:00 hr

0.2 ml/hr

84 hrs

1415 ml

 

 

 

 

 

 

Table 7. 9-volt Alkaline-type batteries used with the CADD-Prizm®pump.

Volume

Duration

Cycle

KVO

Life

Volume

23.5

1:00 hr

5:00 hr

0.2 ml/hr

301 hrs

1229 ml

61

1:00 hr

6:00 hr

0.2 ml/hr

183hrs

1891 ml

125

1:00 hr

6:00 hr

0.2 ml/hr

120 hrs

2520 ml

200

1:00 hr

12:00 hr

0.2 ml/hr

186 hrs

3134 ml

 

 

 

 

 

 

Table 8. 9-volt Lithium-type batteries used with the CADD-Prizm®pump.

TPN Delivery Battery Life

Note: Results are without air detector.

Volume

PeriodTaper UpTaper Down

KVO

Life

Volume

 

 

 

 

 

 

 

2000

10:00 hr

1:00 hr

1:00 hr

5 ml/hr

12 hr

2333 ml

3000

10:00 hr

1:00 hr

1:00 hr

5 ml/hr

8 hr

2497 ml

 

 

 

 

 

 

 

Table 9. 9-volt Alkaline-type batteries used with the CADD-Prizm®pump.

Volume

Period

Taper Up

Taper Down

KVO

Life

Volume

 

 

 

 

 

 

 

2000

10:00 hr

1:00 hr

1:00 hr

5 ml/hr

32 hr

6333 ml

3000

10:00 hr

1:00 hr

1:00 hr

5 ml/hr

19 hr

5833 ml

 

 

 

 

 

 

 

Table 10. 9-volt Lithium-type batteries used with the CADD-Prizm®pump.

Volume

Period

Taper Up

Taper Down

KVO

Life

Volume

 

 

 

 

 

 

 

2000

10:00 hr

1:00 hr

1:00 hr

5 ml/hr

64 hr

12777 ml

3000

10:00 hr

1:00 hr

1:00 hr

5 ml/hr

45 hr

13498 ml

 

 

 

 

 

 

 

Table 11. EPS System used with the CADD-Prizm®pump.