Continuous and PCA Delivery Battery Life (Max Delivery Rate PCA Mode 30 ml/hr)

Note: Results are without air detector.

RateLife

Volume

0.4 ml/hr

120 hrs

48 ml

10 ml/hr

86 hrs

860 ml

30 ml/hr

37 hrs

1110 ml

50 ml/hr

26 hrs

1300 ml

100 ml/hr

13 hrs

1300 ml

200 ml/hr

14 hrs

2800 ml

350 ml/hr

7 hrs

2450 ml

 

 

 

Table 4. 9-volt Alkaline-type batteries used with the CADD-Prizm®pump.

Rate

Life

Volume

0.4 ml/hr

212 hrs

85 ml

10 ml/hr

161 hrs

1610 ml

30 ml/hr

79 hrs

2370 ml

50 ml/hr

60 hrs

3000 ml

100 ml/hr

30 hrs

3000 ml

200 ml/hr

32 hrs

6400 ml

350 ml/hr

17 hrs

5950 ml

 

 

 

Table 5. 9-volt Lithium-type batteries used with the CADD-Prizm®pump.

Rate

Life

Volume

100 ml/hr

64 hrs

6400 ml

200 ml/hr

67 hrs

13400 ml

350 ml/hr

39 hrs

13650 ml

 

 

 

Table 6. EPS System used with the CADD-Prizm®pump.