Pre- | Post | Weight of | Intended | Inaccuracy | Accuracy | Accuracy |
Delivery | Delivery | Amount | Delivery | Volume | Error | Error Percentage |
Weight | Weight | Delivered | Volume |
|
|
|
|
|
|
|
|
|
|
61.1 g | 41.6 g | 19.5 g = 19.5 ml | 20 ml | |||
|
|
|
|
|
| |
|
|
|
|
|
|
|
Table 14. Gravimetric percentage calculation |
|
|
|
|
11.If the accuracy error percentage is greater than ± 6%, repeat the test with a new reservoir. If the pump fails a second time, call Deltec’s Customer Service Department.
EXAMPLE:
Predelivery Weight: | 61.1 g | |
Postdelivery Weight: | - 41.6 g |
|
Weight of Amount Delivered: | 19.5 g | |
| = 19.5 ml | |
Volume of Amount Delivered: | 19.5 ml | |
Intended Delivery Volume: | - 20.0 ml |
|
Inaccuracy Volume: | – 0.5 ml | |
Inaccuracy Volume: | – 0.5 ml | |
Intended Delivery Volume: | ⎟ 20.0 ml |
|
Accuracy Error: | ||
Accuracy Error: |
| |
| x 100.00 |
|
Accuracy Error Percentage: |
Volumetric Accuracy Testing (Option 2)
Description:
A predetermined amount of water is delivered into a collection device such as a burette or graduated cylinder. The amount of water delivered is compared to the amount that the pump should have delivered.
Nominal system accuracy is given in the technical specifications section for the pump. That is, under the test conditions described below, the accuracy of the pump and reservoir will be nominal with a 90% confidence level. The nominal test conditions are as follows: degassed water at 25 ± 5°C without back pressure.
Equipment needed:
•50- or
•50- or
•CADD® extension set with
•A fluid collection device such as a burette or a class A 25 ml capacity graduated cylinder
•40 ml of room temperature water
Procedure:
1.Fill the 50- or
2.Remove any air from the Medication Cassette reservoir by aspirating the air with the syringe. Attach the CADD® extension set with
3.Attach the end of the extension set to the fluid collection device.
4.Attach the Medication Cassette reservoir to the pump. Program the reservoir volume to 20 ml. This is the intended delivery volume. Open all clamps.
5.Program a continuous rate of 0.0 ml/hr and a Demand Dose of 1.0 ml (but do not deliver a Demand Dose). Start the pump and deliver a clinician bolus of
20 ml.
6.When delivery is complete, record the volume of fluid delivered. This is the actual delivery.
7.Find the difference between the volume of the amount delivered and the intended
29