Specifications (Nominal)
General Pump SpecificationsResolution
Medicaton Cassette reservoir or CADD® administration set, 0.050 ml/pump stroke nominal
Size
4.4cm x 10.4 cm x 14.1 cm (1.7 in. x 4.1 in. x 5.6 in.) excluding cassette or other accessories
Weight
568 g (20 oz.) including
Pump Alarms
Low battery power; depleted battery power; external power source low, faulty, depleted; pump stopped; pump fault; low reservoir volume; high delivery pressure; air in line; air detector faulty or detached (only with the use of the optional air detector); air detector port cover detached; delivery too slow; key stuck; cassette detached or unlocked; print failure, upstream occlusion (model 6101 only)
Bolus Volume at Occlusion Alarm Pressure
0.050 ml resolution sets/reservoirs: < 0.25 ml
0.100 ml resolution sets: < 2.0 ml
Power Sources
An internal battery powers the clock. When it is depleted, it cannot reliably maintain the clock time. This battery must be replaced by the manufacturer. The internal battery has an expected life of 5 years.
System* Operating Temperature +2°C to 40°C (36°F to 104°F)
System* Storage Temperature
Power Pack Charging Temperature +10°C to 35°C (50°F to 95°F)
System* Delivery Accuracy
±6% (nominal)
High Pressure Alarm 18 ± 9 psi
Air Detector Alarm
Single bubble greater than 0.100 ml
PCA Delivery Mode SpecificationsReservoir Volume
1 to 9999 or Not In Use; programmable in
1 ml increments, displayed in 0.1 ml increments Default: 1 ml
Units
Milliliters (ml), milligrams (mg), micrograms (mcg) Default: milligrams
Concentration
Mg/ml: 0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 10, 15, ... 95, 100
Mcg/ml: 1, 2, 3, 4, 5, 10, 15, ...95, 100, 200, 300, 400, 500
Default: 1 mg/ml
Continuous Rate
0 to 30 ml/hr (or the mg or mcg equivalent) (See Table 1 for scroll ranges)
Default: 0 mg/hr
Demand Dose
0 to 9.9 ml (or the mg or mcg equivalent) (See Tables 2, 3 and 4 for scroll ranges) Default: 0 mg/hr
Delivery rate (Continuous Rate + Demand Dose): 125 ml/hr nominal
Demand Dose Lockout
5 minutes to 24 hours in the following increments:
1 minute for values between 5 and 20 minutes
5 minutes between 20 minutes and 24 hours Default: 5 minutes
Max Doses Per Hour
1 to 12 doses in 1 dose increments (will also be limited by the Demand Dose Lockout value) Default: 1
Demand Doses Given 0 to 999
Demand Dose Attempts 0 to 999
*System is defined as a