Eizo CG245W user manual Setting the calibration schedule Schedule, Fri/Sat/Sun

Models: CG245W

1 66
Download 66 pages 20.36 Kb
Page 35
Image 35

Setting the calibration schedule [Schedule]

Set the calibration schedule for SelfCalibration.

Attention

• Clock adjustment is necessary for SelfCalibration to operate.

Procedure

1.Choose <Setting> from the SelfCalibration menu, and press .

2.Choose <Schedule> from the Setting menu, and press . The schedule setting menu is displayed.

3.Select the items to change with or , and press .

Setting Item

Setting Range

 

Setting

Starting Time

Off

Set the timing at which SelfCalibration is executed.

 

Power Save

Off

: SelfCalibration is not executed.

 

Immediately

Power Save

: When the time period set in the schedule

 

 

 

has elapsed, it is executed if the mode

 

 

 

initially switches to power save, or if the

 

 

 

power initially switches off.

 

 

Immediately

: When the time period set in the schedule

 

 

 

has elapsed, it is executed immediately.

Setting Type

Type1

Set according to the frequency at which SelfCalibration is

 

Type2

executed.

 

 

Type3

Type1 : Once every few months

 

 

Type2 : Once every few weeks

 

 

Type3 : Once every few hours (monitor usage time)

Interval

Type1 : 1 - 12 months

Set the interval between the completed SelfCalibration and

 

Type2 : 1 - 5 weeks

the next execution. The setting details differ depending on

 

Type3 : 50 - 500 h

the type of <Setting Type> specified.

Week

1st week - 5th week

Set which week to execute SelfCalibration on for the month

 

 

in which calibration is executed.

Day

MON/TUE/WED/THU/

Set which day to execute SelfCalibration on for the week in

 

FRI/SAT/SUN

which calibration is executed.

Time

00:00 - 23:30

Set the time at which SelfCalibration is executed.

*1 Settings that are used frequently can be preset.

4.Adjust using or .

5.Press to exit the adjustment. The calibration schedule is set.

NOTE

The schedule can be set using ColorNavigator. Refer to the ColorNavigator User’s Manual on the CD-ROM.

When the starting time is [Immediately], and the setting type is either [Type1] or [Type2], SelfCalibration is

executed according to schedule even if the power is switched off using

.

When the starting time is “Off”, the SelfCalibration menu shows “Setting Incomplete.” However, SelfCalibration can be executed manually.

Chapter 4 SelfCalibration 35

Page 35
Image 35
Eizo CG245W user manual Setting the calibration schedule Schedule, Fri/Sat/Sun