46 Chapter6Troubleshooting
6-4. Built-in calibration sensor and SelfCalibration problemsProblem Possible cause and remedy
1. The built-in calibrat ion sensor does not
come out/stays out.
• Turnoffthemainpower,andthenturnitonagainafew
minuteslater.
2. SelfCalibrati on cannot be executed. • Checkwhetherthedateandtimearesetcorrectly(see
“Settingmonitordateandtime”(page36)).
• Checkwhetherthemodeisset(see“Setthemodefor
whichSelfCalibrationisexecuted.”(page38)).
• Checkwhetherthecalibrationtargetsaresetcorrectly(see
“SettingAdjustmentTarget”(page39)).
• SettheSelfCalibrationtargetusingColorN avigator.
3. SelfCalibration f ailure • Refertotheerrorcodetable.Ifanerrorcodethatdoe s
notappearintheerrorcodetableisdisplayed,contac t
yourlocalEIZOrepresentative.
Error Code Table
Error Code Description
0014 • ExecuteSelfCalibrationagain.
• AdjustusingColorNavigator.
0034 • Thebuilt-incalibrationsensormaynothavecomeoutduringc alibration,orlightmay
haveenteredthesensorduringmeasurement.
• Turnoffthemainpower,waitforafewminutesbeforeturningthepowerbackon,and
thenexecuteSelfCalibrationagain.
0050 • ExecuteSelfCalibrationagain.
0061 • Thebuilt-incalibrationsensormaynothaveworkedcorrec tly.
• Checkwhetherthereisanyforeignobjectnearthebuilt- incalibrationsensor.
• ExecuteSelfCalibrationagain.