Section 3: Electrical installation | Daniel 3818 LNG Liquid Ultrasonic Meter Installation Manual |
May 2012 | |
|
|
3.6.2Analog input settings
The 3818 LNG Liquid Ultrasonic Flow Meter has the capability to sample analog temperature (Analog Input 1) and pressure (Analog Input 2) with
3.6.3Analog output settings
The 3818 LNG Liquid Ultrasonic Flow Meter provides two
Analog Output 1 (AO1) may used for HART slave communications with the current driven in sink mode (see Appendix A Daniel drawing
Full HART® functionality is provided so that any commercially available HART® transmitter which meets the specifications of the HART® Communications Foundation can be connected to the Daniel Liquid Ultrasonic Flow Meter.
Analog Output 2 (AO2) is
Both Analog outputs have an additional 150 Ohms loop impedance to meet the HART Communications requirements. The 24VDC isolated power supply can be used to power the loop.
3.6.4Digital Input
The 3818 LNG Liquid Ultrasonic Flow Meter provides one digital input that can be used as a general purpose input or used for synchronizing calibration (for flow calibration gating - contact closure). The meter records the volume seen between switch closures. The polarity of the input is configured as normal or inverted polarity.
•polarity is determined by the IsDI1ForCalActiveLow and the gating edge is determined by the IsDI1ForCalStateGated (calibrate edge gated or calibrate state gated)
•calibration is started via an inactive>active state change and stopped via an active>inactive state change.
The digital input must be configured via the Daniel MeterLink Tools>Edit/Compare Configuration page.
52 | Analog input settings |