Daniel 3818 LNG Liquid Ultrasonic Meter Installation Manual | Section 3: Electrical installation | |
May 2012 | ||
|
|
|
3.6.5DHCP server switch settings
The meter can be configured to act as a DHCP server. The DHCP server is enabled/disabled via CPU Module DHCP switch as follows:
Table 3-6 DHCP server switch settings
CPU Module switch | DHCP server disabled | DHCP server enabled |
|
|
|
DHCP | OFF | ON |
|
|
|
3.6.6Configuration protect switch settings
The meter’s configuration parameters and firmware can be protected against changes via CPU Module Write PROT. switch as follows:
Table 3-7 Configuration protect switch settings
CPU Module switch | Configuration protected | Configuration unprotected |
|
|
|
WRITE PROT. | ON | OFF |
|
|
|
A complete list of
3.6.7External power source connection and fuse
Located inside the Transmitter Electronics Enclosure is a connector for a
Figure 3-7 CPU Module power source connections
C.A.B.
A. Power In connector (main power)
B. 24V Loop power
C. 2 Ampere fuse (used for the main power input)