Daniel 3818 LNG Liquid Ultrasonic Meter Installation Manual

Section 4: Configuring a 3818 LNG Meter

3-9000-771 Rev A

May 2012

 

 

 

5.Wire Analog Input 1 (AI1) and Analog Output 1 (AO1) as shown in Figure 4-22.

Figure 4-22 Field Communicator wiring diagram for the 3818 LNG Meter

6.Use the leads provided with the Field Communicator to connect to your device.

7.Press and hold the Power button on the Field Communicator until the green light blinks.

8.Use the touch screen on the Field Communicator, the keypad or use the stylus to navigate through the device menus.

9.Refer to the Menu tree in Section D.1.1 of the Daniel HART Field Device Specification manual (P/N 3-9000-762) for the device fast key sequences. Included in the menu tree are:

Diagram Page 1 - 3810 Series Root Menu; Overview, Configure>Manual Setup

Diagram Page 2 - Configure>Manual Setup (continued) and Alerts Setup

Diagram Page 3 - Service Tools>Alerts and Variables

Diagram Page 4 - Service Tools>Variables (continued), Service Tools>Trends, and Service Tools>Maintenance

10.If you encounter problems, refer to the contact information on the back cover of this manual or the contacts included in the Field Communicator Users Manual.

Using a Field Communicator to configure the meter