Daniel 3818 LNG Liquid Ultrasonic Meter Installation Manual | Section 4: Configuring a 3818 LNG Meter | |
May 2012 | ||
|
|
|
Use the Manual Setup wizard to configure the meter’s parameters. See Figure
Procedure
1.If installed, remove security wires from the endcap and the Bracket/Cover hex head bolts that secures the Base Enclosure.
1.Disable the Write Protect switch in the CPU Module to write any of the following configuration parameters to your meter.
2.Click the Device Variables Mapping tab. The Primary and Secondary variables are read only and are configured for Uncorrected Flow Rate. The Third and Fourth variable configuration choices include Pressure and Temperature.
3.Click the Units tab (see AMS Device Manager - Guided Setup, Step 1).
4.Click the Analog Output 1 (HART) tab (see AMS Device Manager - Guided Setup, Step 2a.).
5.Click the Analog Output 2 tab. Follow the configuration instructions in the AMS Device Manager - Guided Setup, Step 2b. The read only Secondary variable Content, Uncorrected Flow Rate, displays. Use the
Using AMS Device Manager to configure the meter | 69 |