Instruction Manual
Model 755A
a. | General1 |
|
| Operating Range ........................... | 0.00% to 100.0% oxygen |
| Recorder Range ............................ | Selectable for 0% to 100% oxygen or for any desired span of 1%, |
|
| 2%, 5%, 10%, 20% or 100% oxygen within the overall range. |
| Response Time ............................. | (90% of fullscale) recorder output factory set for 20 seconds; |
|
| adjustable from 5 to 25 seconds. |
| Reproducibility (Digital Display)..... | ±0.01% Oxygen ±2 counts. |
| Ambient Temperature Limits |
|
| Maximum ............................... | 49°C (120°F) EXCEPT 38°C (100°F) for 99% to 100% oxygen. |
| Minimum ................................ | |
| Zero and Span Drift2...................... | Within ±1% of fullscale (±2% of fullscale for 99% to 100% range) |
|
| per 24 hours, provided that ambient temperature does not change |
|
| by more than 11.1°C (20°F). |
|
| ±2.5% of fullscale per 24 hours with ambient temperature change |
| Barometric Pressure | over entire range. |
|
| |
| Compensation ....................... | Oxygen readout automatically corrected to within ±1% of fullscale |
|
| for barometric pressure variations within ±3% of target value and |
|
| within ±2% of fullscale for barometric pressure variations within |
|
| ±5% of target value. |
|
| The target may be set anywhere within range of |
|
| psig |
|
| Exhaust vented to atmosphere. |
b. | Sample |
|
| Dryness ......................................... | Sample dewpoint below 43°C (110°F), sample free of entrained |
|
| liquids. |
| Temperature Limits |
|
| Maximum ............................... | 66°C (150°F) |
| Minimum ................................ | 10°C (50°F) |
| Operating Pressure |
|
| Maximum ............................... | 69 kPa (10 psig). |
| Minimum ................................ | |
| Flow Rate 3 |
|
| Maximum ............................... | 500 cc/min |
| Minimum ................................ | 50 cc/min |
| Recommended ...................... | 250 ±20 cc/min |
| Materials in Contact with |
|
| Sample Gas........................... | 316 stainless steel, glass, titanium, Paliney No. 7, epoxy resin, |
|
|
1Performance specifications based on recorder output.
2Zero and span drift specifications based on following conditions: Operating pressure constant; ambient temperature change from initial calibration temperature, less than 11.1 Celsius degrees (20 Fahrenheit degrees); deviation from set flow held to within ±10% or ±20 cc/min, whichever is smaller.
3Deviation from set flow would be held to within ±10% or ±20 cc/min, whichever is smaller. If so, zero and span drift will be within specifications, provided that operating temperature remains constant.
Rosemount Analytical Inc. A Division of Emerson Process Management |