Instruction Manual

245364-V May 2002

Model 755A

1-7 SPECIFICATIONS

a.

General1

 

 

Operating Range ...........................

0.00% to 100.0% oxygen

 

Recorder Range ............................

Selectable for 0% to 100% oxygen or for any desired span of 1%,

 

 

2%, 5%, 10%, 20% or 100% oxygen within the overall range.

 

Response Time .............................

(90% of fullscale) recorder output factory set for 20 seconds;

 

 

adjustable from 5 to 25 seconds.

 

Reproducibility (Digital Display).....

±0.01% Oxygen ±2 counts.

 

Ambient Temperature Limits

 

 

Maximum ...............................

49°C (120°F) EXCEPT 38°C (100°F) for 99% to 100% oxygen.

 

Minimum ................................

-7°C (20°F) EXCEPT 4°C (40°F) for 99% to 100% oxygen.

 

Zero and Span Drift2......................

Within ±1% of fullscale (±2% of fullscale for 99% to 100% range)

 

 

per 24 hours, provided that ambient temperature does not change

 

 

by more than 11.1°C (20°F).

 

 

±2.5% of fullscale per 24 hours with ambient temperature change

 

Barometric Pressure

over entire range.

 

 

 

Compensation .......................

Oxygen readout automatically corrected to within ±1% of fullscale

 

 

for barometric pressure variations within ±3% of target value and

 

 

within ±2% of fullscale for barometric pressure variations within

 

 

±5% of target value.

 

 

The target may be set anywhere within range of -2.7 to 3.3 psig ±3

 

 

psig (-18.6 to 22.8 kPa ±21 kPa).

 

 

Exhaust vented to atmosphere.

b.

Sample

 

 

Dryness .........................................

Sample dewpoint below 43°C (110°F), sample free of entrained

 

 

liquids.

 

Temperature Limits

 

 

Maximum ...............................

66°C (150°F)

 

Minimum ................................

10°C (50°F)

 

Operating Pressure

 

 

Maximum ...............................

69 kPa (10 psig).

 

Minimum ................................

-13.1 kPa (-1.9 psig)

 

Flow Rate 3

 

 

Maximum ...............................

500 cc/min

 

Minimum ................................

50 cc/min

 

Recommended ......................

250 ±20 cc/min

 

Materials in Contact with

 

 

Sample Gas...........................

316 stainless steel, glass, titanium, Paliney No. 7, epoxy resin,

 

 

Viton-A, platinum, nickel.

1Performance specifications based on recorder output.

2Zero and span drift specifications based on following conditions: Operating pressure constant; ambient temperature change from initial calibration temperature, less than 11.1 Celsius degrees (20 Fahrenheit degrees); deviation from set flow held to within ±10% or ±20 cc/min, whichever is smaller.

3Deviation from set flow would be held to within ±10% or ±20 cc/min, whichever is smaller. If so, zero and span drift will be within specifications, provided that operating temperature remains constant.

1-4 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management