Instruction Manual
Model 755A
d.Isolated Current Output (Optional)
1.Verify that the Current Output Board appropriate to desired output is properly in place. See Figure
2.On TB2, Figure
3.Connect free end of output cable to input terminals of recorder or other current actuated device, making sure that polarity is correct. If two or more
Total resistance of all output devices and associated interconnection cable must not exceed 850 ohms.
Current and voltage outputs may be utilized simultaneously, if desired.
+ | + |
mA - | - Recorder |
755A
Analyzer+
- Controller
+
- Remote Indicator
Figure
e.Output Connections for Dual Alarm Option
If so ordered, the analyzer is factory- equipped with alarm output. Alternatively, the alarm feature is obtainable by subsequent installation of the 618083 Alarm Relay Kit.
f.Alarm Output Connections
The alarm output provides two sets of relay contacts for actuation of alarm and/or process control functions. Leads from the
Note the following recommendations:
1.A line fuse should be installed in the line between the
power supply and the alarm relay terminals on the Alarm Relay Assembly.
2.If the alarm contacts are connected to any device that produces radio frequency interference (RFI), it should be
3.If possible, the analyzer should operate on a different AC power source to avoid RFI.
4.Do not allow internal cable service loop to touch the detector assembly or associated inlet and outlet tubing. This precaution ensures against possible transmission of mechanical vibration through the cable to the detector, which can cause loss of accuracy.
Rosemount Analytical Inc. A Division of Emerson Process Management |