Instruction Manual
Model 755A
•Desired oxygen range for recorder output: 99% to 100%.
•Front panel TEST switch is set to position 4, and Recorder Oxygen Span Jumper is placed in 100% oxygen (1X gain) position.
•Required zero suppression is 99% oxygen, thus Recorder Zero Suppression Jumper is set in 80% position, and Recorder Zero Suppression Adj4ustments R41 and R104 are set for a reading of 99.00 volts on the digital display.
•Recorder Oxygen Span Jumper is returned to 1% oxygen (100X gain) position, normal span setting for 99% to 100% oxygen range. R88 may be used for fine span adjustment.
detector compartment with fingers; wait 30 seconds;
When
After analyzer
Calibration for oxygen readout consists of establishing a downscale and upscale point.
•Set front panel TEST switch to position 1.
WARNING
POSSIBLE EXPLOSION HAZARD
If gases are introduced into this analyzer, the sample containment system must be carefully
Internal leakage of sample resulting from failure to observe these precautions could result in an explosion causing death, personal injury, or property damage.
Pass suitable
a.Calibration Using Digital Readout for Oxygen Readout
The digital display covers the full range of 0.00% to 100.005 oxygen and thus will normally be used as the readout device during calibration. If so, almost any
b.Calibration Using Recorder for Oxygen Readout
In some applications the recorder readout may be the important display and may thus be used during calibration. If so, the
1.A
Rosemount Analytical Inc. A Division of Emerson Process Management |