Instruction Manual
6 Digit LCD Display
Model 755A
ZERO |
|
SPAN |
|
PRESS CAL1 | PRESS CAL 2 |
NORM | REC OFFSET |
ZERO Adjust | SPAN Adjust |
TEST Switch | Rosemount Analytical |
|
M o d e l 7 5 5 A
O x y g e n A n a l y z e r
| CONTROL | FUNCTION |
| |
Digital Display (LCD) | Readout of sample oxygen content (0.00% to 100.00%) or selected test function, | |||
depending on position of TEST switch. |
| |||
|
|
| ||
TEST Switch (S1) | Selects variable desired for readout on digital display. |
| ||
| Switch Position | Readout on Digital Display | Designation | |
1 | Oxygen content of sample (0.00% to 100.00%) | NORM | ||
|
|
|
| |
| 2 | Pressure/Voltage CAL 1 | CAL 1 | |
3 | Pressure/Voltage CAL 2 | CAL 2 | ||
|
|
|
| |
4 | Zero suppression applied to analog output circuit | ZERO SUPPRESSION | ||
(0.00 to 10.00 VDC) | ||||
|
|
| ||
|
| Used to establish downscale calibration point on digital display or recorder chart. | ||
ZERO Control (R13) | With suitable downscale standard gas flowing through the analyzer, the ZERO | |||
|
| Control is adjusted for appropriate reading on display. |
| |
|
| Used to establish upscale calibration point on digital display or recorder chart. With | ||
SPAN Control (R16) | suitable upscale standard gas flowing through the analyzer, the SPAN Control is | |||
|
| adjusted for appropriate reading on display or recorder. |
| |
|
| Figure |
|
Operation | Rosemount Analytical Inc. A Division of Emerson Process Management |