Model 755A
Instruction Manual
Pressure |
| F1 | Compound Range |
| |
|
| Pressure Gauge or |
| ||
Gauge |
|
| Mercury Manometer |
| |
|
|
|
| Vacuum | |
PR1 |
|
|
| Pump | |
|
| Sample |
| ||
Sample In |
|
|
| ||
| 250 | In 755A |
| ||
10 psig | V1 | V3 | |||
cc/min |
| ||||
|
|
| |||
|
|
| V2 | V4 |
Figure
Standard Procedure: Calibration by Pressure Decrease
1.Verify that the Pressure Compensation Option is selected, i.e., Jumper inserted between E1 and E3 on the Control Board.
2.Connect a compound range pressure gauge (or mercury manometer), pressure gauge, pressure regulator, flowmeter, needle valves and vacuum pump as shown in Figure
3.Open V2, set PR1 to approximately 10 psig (69 kPa) and adjust V1 for a flow rate of 250 cc/min. into the sample inlet line. Compound range pressure gauge (or mercury manometer) should indicate close to atmospheric pressure (0.00).
4.Set TEST switch to PRESS CAL 1 and adjust CAL 1 potentiometer R98 (see Table
5.Set TEST switch to NORM and note display reading.
6.Start vacuum pump.
7.Close V2 and adjust V3 and V4 to decrease the pressure on the sample exhaust by about 3%. (- 12.00 inches of water;
8.Adjust CAL 2 potentiometer R99 (see Table
9.Repeat steps 3 through 7 as many times as necessary to ensure repeatability of compensation calibration.
NOTE
Although not mandatory, it is desirable, in order to obtain maximum accuracy, that the oxygen concentration of the gas used during calibration be close to that of the process stream which is to be analyzed.
Rosemount Analytical Inc. A Division of Emerson Process Management | Operation |